世界中医药
文章摘要
引用本文:龙华晴1,2,吴人照1,吴焕淦3,寿张根4,赵蕾1.生物信息反馈红外治疗仪对局部与远端皮肤温度影响的临床观察[J].世界中医药,2016,(12):.  
生物信息反馈红外治疗仪对局部与远端皮肤温度影响的临床观察
Clinical Observation of Effect of Local and Distal Skin Temperature by Biological Information Feedback Infrared Therapeutic Apparatus
投稿时间:2016-12-07  
DOI:10.3969/j.issn.1673-7202.2016.12.004
中文关键词:  红外线  脉动    皮肤温度  生物信息反馈红外治疗仪
English Keywords:Infrared ray  Pulse  Moxibustion  Skin temperature  Biological information feed-back infrared therapy apparatus
基金项目:2011年国家科技部火炬计划(编号:2011GH041021);国家重点基础研究发展计划(“973”计划)项目(编号:2009CB522900;2015CB554501)
作者单位
龙华晴1,2,吴人照1,吴焕淦3,寿张根4,赵蕾1 1 浙江省中医药研究院杭州310007 2 浙江中医药大学杭州310051 3 上海中医药大学上海市针灸经络研究所上海200030 4 杭州大力神医疗器械有限公司杭州311200 
摘要点击次数: 852
全文下载次数: 774
中文摘要:
      目的:采用生物信息反馈红外治疗仪,观察指尖微动脉搏动节律同步脉动照射和连续照射两种不同照射方式对局部皮肤温度和远端劳宫穴皮肤温度的影响,探讨随指尖微动脉搏动节律行同步脉动照射方式是否优于连续照射方式。方法:将纳入照射研究的60人次随机分为2组:同步脉动组、连续照射组,每组30人次,观察不同照射方式对局部脐周和右手肘关节及远端右手劳宫穴每隔1 min的皮肤温度、指脉微循环脉图高度、心率的情况。结果:脐周照射同步脉动组与连续照射组照射前、后局部皮肤温度和右手劳宫穴皮肤温度均显著升高(P<0.01),但照射前后局部皮肤温度升高同步脉动组为(3.53±1.39)℃,显著低于连续照射组的(4.62±1.38)℃(P<0.01)。15 min照射前、后右手劳宫穴皮肤温度升高同步脉动组为(1.45±1.07)℃,高于连续照射组的(1.08±1.00)℃,但差异无统计学意义(P>0.05)。肘关节照射后结果与脐周照射相似。结论:体表微循环指脉同步脉动照射与连续照射相比,体表微循环指脉同步脉动照射的局部皮肤温度较低,为实现充足灸量创造了条件,远端劳宫穴皮肤温度较高,更能促进人体远端微循环。说明生物信息反馈红外治疗仪大力神雀啄灸的同步脉动照射方式比连续照射方式的在能量利用方面更有优势。
English Summary:
      To observe the effect of local skin temperature and the remote skin temperature on PC8 acupoint under two different irradiations of synchronous pulse beam (the stimulus frequency was associated with the pulse of figure provided by biological information feedback infrared therapeutic apparatus)and continuous illumination on surface microcirculation (finger pulse), and to discuss whether the energy utilization of synchronous pulse radiation on surface microcirculation (finger pulse) is better than the continuous illumination. Methods:A total of 60 patients were randomly divided into two groups:synchronous pulse group and continuous irradiation group, 30 patients in each group, and then we observed the effects of different exposure modes on skin temperature, finger pulse microcirculation, and heart rate at 1 minute intervals for local (peripheral umbilicus and right elbow) and distal right palm (PC8). Results:The skin temperature and palmar skin temperature of the two groups rose significantly (P<0.01), but the temperature of local skin in the synchronized pulse group was (3.53±1.39)℃, which was significantly lower than that in the continuous irradiation group (4.62±1.38)℃, P<0.01). After 15 minutes of irradiation the skin temperature of the right hand palm (PC8) was (1.45±1.07)℃, which was higher than that of the continuous irradiation group (1.08±1.00)℃, but there was no significant difference (P>0.05). Elbow irradiation result was similar to periumbilical irradiation result. Conclusion:By contrast, surface skin microcirculation (finger pulse) synchronous pulsation of the local skin temperature is lower, that it created the conditions to achieve “adequate moxibustion amount”. The skin temperature was higher at the distal palm, and that can promote the human body microcirculation. The experimental results show that the synchrotron pulsed irradiation method of bioinformation feedback infrared therapeutic apparatus is superior to the continuous irradiation method.
查看全文  查看/发表评论  下载PDF阅读器